Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: GE Healthcare Panda iRes Warmers integrated with Nellcor SpO2

22 Apr 2014

The Australia Therapeutic Goods Administration (TGA) posted a medical device safety alert concerning Panda iRes Warmers integrated with Nellcor SpO2, manufactured by GE Healthcare.

There may be a loss of internal communication that manages the user-adjusted Nellcor SpO2 alarm settings of the affected Giraffe and Panda iRes Warmer systems with Nellcor SpO2 listed above. If the issue exists, the alarms may not activate as expected, which can result in false positive or false negative saturation and pulse rate alarm notifications to the caregiver. The patient's oxygen saturation and pulse rate values are accurate, as displayed; however, the displayed user-set alarm limits may be different than the actual alarm limits used for alarm activation. All other clinical functionality of the warmer is unaffected. This alarm failure is caused by a software issue that does not detect a loss of internal communication that manages the user-adjusted Nellcor SpO2 alarm settings.

The manufacturer will correct all affected systems with a software revision. Customers may continue to use the system provided they follow the GE recommended actions as follows:

For details, please refer to TGA website:
http://www.tga.gov.au/SARA/arn-detail.aspx?k=RC-2014-RN-00450-1

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 22 April 2014

Back