Medical Device Safety Alert: Endoscopic Retrograde Cholangiopancreatography (ERCP) endoscopes

23 Feb 2015

The United States Food and Drug Administration (FDA) has issued an alert that the complex design of ERCP endoscopes may impede effective reprocessing. Reprocessing is a detailed, multistep process to clean and disinfect or sterilize reusable devices. This alert affects ERCP endoscopes supplied by all manufacturers.

Recent medical publications and adverse event reports associate multidrug-resistant bacterial infections in patients who have undergone ERCP with reprocessed endoscopes even when manufacturer reprocessing instructions are followed correctly. From January 2013 through December 2014, the FDA received 75 medical device reports encompassing approximately 135 patients in the United States relating to possible microbial transmission from reprocessed endoscopes. Meticulously cleaning endoscopes prior to high-level disinfection should reduce the risk of transmitting infection, but may not entirely eliminate it.

The FDA is advising health care professionals to

Posted on 23 February 2015