Medical Device Safety Alert: Brainlab ExacTrac v. 6.x – Cone Beam Computed Tomography (CBCT) Import & Alignment Software

23 Apr 2015

Medical device manufacturer, Brainlab AG, has issued a medical device safety alert concerning its ExacTrac v. 6.x – Cone Beam Computed Tomography (CBCT) Import & Alignment Software [Software version: ExacTrac v. 6.x (v. 6.0.0, 6.0.1, 6.0.2, 6.0.3, 6.0.4, 6.0.5; and v. 6.1.0), in combination with Varian C-Series linac only].

According to the manufacturer, an anomaly was detected in ExacTrac v.6.x during phantom based acceptance testing. The phenomenon occurs under specific circumstances when using the ExacTrac CBCT module with Varian CBCT images acquired with a Varian C-Series linear accelerator (linac) at a couch angle other than 0.0°.

The manufacturer advises users to take the following actions:

The manufacturer will provide a software update to affected customers.

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.