Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Terumo BCT Spectra Optia Apheresis Systems

01 Sep 2015

Medical device manufacturer, Terumo BCT, has issued a medical device safety alert concerning its Spectra Optia Apheresis Systems.

The manufacturer has received reports of data entry errors that cause the system to calculate a patient total blood volume (TBV) that is abnormal for that specific patient. The error can occur under the following circumstances:

If the operator does not enter correct data, the patient may receive excess anticoagulant (AC) or the outcome of the procedure may be affected. This issue can have the greatest effect on small or compromised patients for whom an abnormally high TBV calculation could lead to hypocalcemia.

The manufacturer instructs affected users to review and confirm the data entered on the systems screens before connecting the patient to the system and throughout the procedure. The manufacturer will also provide software upgrade to mitigate the potential for data entry errors.

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 1 September 2015

Back