Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Iris International iChemVELOCITY Urine Chemistry Strips

26 Jan 2016

Medical device manufacturer, Iris International, has issued a medical device safety alert concerning its iChemVELOCITY Automated Urine Chemistry System using iChemVELOCITY Urine Chemistry Strips [Reference number: 800-7212; 800-7212-001; 800-7204; 800-7204-FJR].

The manufacturer has become aware of loose or missing analyte pads being found in several locations (strip vial, strip provider module, strip conveyor system or waste container) that are undetectable by the iChemVELOCITY System. If affected strips are used to assay patient samples, the issue has the potential to cause erroneous pH results and false negative results for the following tests: nitrite, glucose, blood, bilirubin, urobilinogen, protein, leukocyte esterase, ketones, and ascorbic acid. There is no impact to the measurement of colour, clarity and specific gravity.

Affected users are advised to take the following actions:-

Product replacement is on-going.

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 26 January 2016

Back