Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: BioMerieux Etest COLISTIN CO256 WW

02 Aug 2016

Medical device manufacturer, BioMerieux, has issued a field safety notice concerning its Etest COLISTIN CO256 WW. The affected product reference numbers are 537300 and 537308.

Based on scientific publications (including EUCAST warning letter on their website) on colistin gradient and disk diffusion tests, the manufacturer was alerted about a potential underestimation of the minimum inhibitory concentrations (MIC) values for Enterobacteriaceae and Acinetobacter, the manufacturer initiated a complaint investigation to confirm product performance and determine root cause. While the internal and external investigations are still ongoing, to date and based on data available, the following have been identified:

The manufacturer requests users to take the following actions:

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the products, please contact your supplier for necessary actions.

Posted on 2 August 2016

Back