Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Boston Scientific Cosman Medical Nitinol RF (TCN) Reusable Electrodes

07 Jul 2017

Medical device manufacturer, Boston Scientific, has issued a medical device safety alert concerning its Cosman Medical Nitinol RF (TCN) Reusable Electrodes. The TCN Electrodes were manufactured by Cosman Medical, a wholly-owned subsidiary of Boston Scientific, and affected units are labeled as Cosman. The affected products are TCN-5, TCN-15-3M, TCNK-5-R, TCNK-15-C, TCN-5-3M, TCN-20, TCNK-10, TCNK-15-R, TCN-10, TCN-20-3M, TCNK-10-C, TCNK-20, TCN-10-3M, TCNK-5, TCNK-10-R, TCNK-20-C, TCN-15, TCNK-5-C, TCNK-15 and TCNK-20-R.

According to the manufacturer, there have been complaints that after multiple reprocessing cycles, the epoxy resin which holds the TCN Electrode in the hub can exhibit signs of damage. In some cases, this damage may result in the inability to fully remove blood and/or tissue residuals prior to cleaning and re-sterilizing the device. The failure to completely remove any blood and/or tissue from the TCN Electrodes prior to re-sterilization creates a potential risk of infection and spread of blood borne pathogens to patients. The manufacturer has not received any reports of patient injury related to this issue. Cosman stainless steel (CSK) and Disposable (TCD/CU/CUR) Electrodes are not affected by this action.

The manufacturer advises users to take the following actions:

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 7 July 2017

Back