Medical Device Safety Alert: DePuy DELTA XTEND Reverse Shoulder System Modular Centered and Eccentric Epiphysis Implants

15 Sep 2017

Medical device manufacturer, DePuy, has issued a medical device safety alert concerning its DELTA XTEND Reverse Shoulder System Modular Centered and Eccentric Epiphysis Implants. The affected devices are identified as the following:-

The manufacturer's investigation determined that the affected lots produced in a specific machine were manufactured in a manner in which the post's diameter is out-of-specification. Further investigation determined that the issue is related to the circularity of the cylinder. The circularity issue may cause the epiphysis to interfere with the stem and the two implants (epiphysis and stem) may not be able to be assembled.

According to the manufacturer, there are two potential instances in which the patient could be affected:

According to the local supplier, the affected products are not distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.