Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Abbott Diagnostic Instrument Manager

02 Oct 2018

Medical device manufacturer, Abbott Diagnostics, has issued a medical device safety alert concerning its Instrument Manager [List Number: 03P23; Serial Numbers: 340534, 344571, 340456, 341105, 343400].

The manufacturer has received the information from Data Innovations, Manufacturer of Instrument Manager regarding QC results transmission between Instrument Manager (IM) and Alinity ci interface.

According to the manufacturer, when Alinity ci instrument transmits result from a Quality Control (QC) Sample ID (Q^Control in SPM segment), the IM Alinity ci driver (v8.00.0001) does not parse the results, resulting in QC data not going into IM. This happens when the driver configuration "Hold Results Until Complete" feature is enabled. All affected users using the Instrument Manager software, List Number 03P23, with Alinity ciā€Series interface driver name: abdicqii, v8.00.0001 may be impacted. This issue was identified via a complaint.

The affected users are advised to take the following workaround instructions.

The manufacturer's representative will be contacting the affected users to apply the updated (v8.00.0002) version of Instrument Manager driver to correct this issue.

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the products, please contact your supplier for necessary actions.

Posted on 2 October 2018

Back