Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Abbott Laboratories ARCHITECT Reaction Vessels

06 Dec 2019

Medical device manufacturer, Abbott Laboratories, has issued a medical device safety alert concerning its ARCHITECT Reaction Vessels [List Number: 7C15-03; Multiple Lot Numbers].

The manufacturer observed an increase in complaints due to static discharge in the ARCHITECT Reaction Vessel (RV) on the ARCHITECT i2000, i2000SR, and i1000SR at the optics reader. The static discharge can result in Error Code 1007 (Unable to process test, activated read failure) or potentially falsely elevated/positive results when using the ARCHITECT RV lots.

According to the manufacturer, static discharge impacts direct assays as the lower relative light unit (RLU) counts for direct assays could be impacted by the additional RLUs introduced by these static discharge spikes. There is potential impact to patient results if the static discharge occurs without generating an error code.

Affected users are instructed to take the following actions:-

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 6 December 2019

Back