Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Roche Elecsys CA 19-9 with reagent lot 416245 on cobas e 801

24 Jan 2020

Medical device manufacturer, Roche, has issued a medical device safety alert concerning its Elecsys CA 19-9 with reagent lot 416245 on cobas e 801 [Production Identifier: 07027028 190].

The manufacturer recently received an increased number of complaints concerning non-reproducible elevated results for the affected product.

The issue appears as follows:

According to the manufacturer, the issue can lead to non-reproducible elevated CA 19-9 results and therefore affect clinical interpretation.

The manufacturer's investigations revealed that the occurrence of non-reproducible falsely elevated results is related to a contamination with magnetic/paramagnetic particles (no beads) that occurred during the filling process for cobas e 801 only. Thus, reagents filled for cobas e 411/e 601/e 602 are unaffected.

The manufacturer has already taken immediate corrections and its investigations are currently ongoing to:

Affected users are advised to take the following actions:

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 24 January 2020

Back