Medical Device Safety Alert: Baxter Prismaflex Control Unit

13 Feb 2020

Medical device manufacturer, Baxter, has issued a medical device safety alert concerning its Prismaflex Control Unit [Product Code: 106913, 107493, 103082, 113874, 114489, 114870, 966052; Serial Number: All].

The manufacturer informs important safety information regarding the use of connectors with the Prismaflex system. The use of connectors may prevent a secure connection between the return line and the patient's blood access device. To ensure a proper connection, users must follow the Prismaflex Operator's Manual, which lists the following warnings:

WARNING:

WARNING:

The manufacturer cannot guarantee connectors will establish and maintain secure connections with Prismaflex sets. Additionally, use of connecting devices with the Prismaflex Control Unit could interfere with the ability of the Prismaflex Control Unit to accurately detect pressure drops in the blood circuit. As a result, vascular access disconnects may go undetected, leading to clinically significant blood loss and fatal exsanguination. There has been no reports of product complaint and adverse event associated with this issue in Hong Kong.

Operators may continue to safely use the Prismaflex control units according to the warnings and cautions in the Operator's Manual.

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.