Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Axis-Shield Diagnostics Alere NT-proBNP for ARCHITECT and Alinity i Calibrators and Controls

20 Mar 2020

Medical device manufacturer, Axis-Shield Diagnostics Ltd, has issued a medical device safety alert concerning its Alere NT-proBNP. The affected products are identified as follow:

The manufacturer has identified a time dependent, stability drift in results returned from NT-proBNP testing on both Abbott ARCHITECT and Alinity i systems. The manufacturer is to inform users of a product recall for the affected products for use on the ARCHITECT or Alinity i systems. All of the lots listed demonstrate higher than expected quality control variation and / or patient sample returns when testing on the ARCHITECT and Alinity i systems.

Internal testing has confirmed this is due to an unacceptable stability drift in quality control values, when compared to the initial market release values of the calibrator lots listed. If Calibrators and Controls with the same assigned expiry date are used together the Control values may not be elevated as the material may have similar stability drift. This issue may generate falsely elevated patient results that may go undetected if quality controls do not detect the shift.

According to the manufacturer, affected users are advised to do the following actions:

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 20 March 2020

Back