Medical Device Safety Alert: Beckman Coulter IMMAGE Immunochemistry System APB Apolipoprotein B Reagent, IMMAGE Immunochemistry System LPAX Lipoprotein (a) Reagent, IMMAGE Immunochemistry System PAB Prealbumin Reagent

01 Mar 2021

Medical device manufacturer, Beckman Coulter, has issued a medical device safety alert concerning the following products:

The manufacturer has reported that the following chemistries fail to meet Instructions for Use or IFU lipid interference claims at the following levels:

According to the manufacturer, there is a remote probability of false high results for APB and PAB, and a false low result for LPAX. However, results for all analytes should still fall within total allowable error.

The manufacturer advised that IFUs are located on the Beckman Coulter website's Technical Documents section and affected users should follow the advice on "Resolution" of the notification letter as the information supersedes previous claims and serves as current labeling until the revised IFUs are released.

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the products, please contact your supplier for necessary actions.