Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: bioMérieux MYLA SOFTWARE

08 Sep 2021

Medical device manufacturer, bioMérieux, issued a medical device safety alert concerning the following devices:

Reference Product Name Version
423434 MYLA SOFTWARE V4.8 ; V4.8.1 ; V4.8.2
423270 MYLA SOFTWARE V4.7.1, V4.7
423197 MYLA SERVER DL360 GEN10 NA
423445 MYLA PC XPRO CENTRAL UNIT NA
421993 MYLA VIRTUAL MACHINE NA

Following one customer complaint, the manufacturer confirmed that a MYLA software anomaly can result in a bottle maximum test time (MTT) being reset to the default MTT. The bottle may then be unloaded before the desired incubation time resulting a false negative result.

According to the manufacturer, for this anomaly to occur. The following conditions must be met:

If one or more conditions are not met, the manufacturer determined that user's laboratory is not at risk for this anomaly. However, if all conditions are met, users are advised to contact manufacturer. For tests previously performed using MYLA and VIRTUO, users are advised to identify any possible false negative result that may have occurred, to analyse the related risks and to determine appropriate actions.

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the products, please contact your supplier for necessary action.

Posted on 8 Sept 2021

Back