Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Thermo Fisher Scientific Oxoid Antimicrobial Susceptibility Testing Disc CPD10 (Cefpodoxime)

06 Feb 2013

Medical device manufacturer, Thermo Fisher Scientific issued a medical device safety alert concerning Oxoid Antimicrobial Susceptibility Testing Disc CPD10 (Cefpodoxime) with model number CT1612B and lot number 1275353.

According to the manufacturer, a technical investigation has indicated that the affected lot may contain individual discs that are not impregnated with antibiotic. Continued use of this lot could result in false indication of resistance to Cefpodoxime. After risk assessment, the manufacturer believes the clinical risk of a blank cefpodoxime disc is low.

The manufacturer advised the affected customers to:

The manufacturer confirmed that the affected products have been distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 06 February 2013

Back