Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Philips HeartStart MRx Monitor/Defibrillators

14 Jun 2013

Medical device manufacturer, Philips Healthcare, has issued a medical device safety alert concerning certain models of HeartStart MRx Monitor/Defibrillators [Model no.: M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5 and M3536M6]. The affected units include those manufactured by Philips between February 2004 and February 2013, and shipped worldwide with serial numbers within the ranges US00100100 and US00565942.

A Philips HeartStart MRx Monitor/Defibrillator operating on battery power only that is used in-hospital or in pre-hospital environments may shutdown without warning if exposed to elevated levels of electromagnetic interference from other radio frequency (RF) energy sources and continuous repetitive power line bursts. Examples of other equipment that can cause interference are other medical devices, cellular products, information technology equipment and radio/television transmissions.

The manufacturer provides the following information and advice

In addition, the manufacturer will provide an IFU Addendum titled "HeartStart MRx Electromagnetic Compatibility" for all customers with affected units.

According to the local supplier, the affected products were distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 14 June 2013

Back