Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: GE Healthcare Tec 6 and Tec 6 Plus Vaporizers

13 Aug 2013

Medical device manufacturer, GE Healthcare issued a medical device safety alert concerning all Tec 6 and Tec 6 Plus Vaporizers.

According to the GE Healthcare, the user manual pre-operative check, low pressure leak test at 1% dial setting may not detect the full range of leaks from seal wear degradation in the vaporizers. Moreover, the integrity of the vaporizer seal may be affected with Tec 6 or Tec 6 Plus Vaporizer remanufactured/ manufactured between 12 Aug 2005 and 3 Dec 2012, due to wear degradation.

For any Tec 6 or Tec 6 Plus Vaporizer remanufactured / manufactured between 12 Aug 2005 and 3 Dec 2012 (AETJ29001 to AESR49001), a GE Healthcare Service Representative will schedule a repair to update the vaporizer seal. The leak test failures will get first priority for scheduling repair/replacement

The manufacturer advises that the addendum with the field safety notice distributed to the affected users had the 12% low pressure leak test which will aid in determining the integrity of the seal. The manufacturer also provides the following recommendations to the affected users:

According to the local supplier, the affected products have been distributed to Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 13 August 2013

Back