Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Invacare Power Wheelchairs

07 Oct 2013

The Australia Therapeutic Goods Administration (TGA) has posted a medical device safety alert concerning power wheelchairs manufactured by Invacare Corporation. The following products are affected:

Invacare has identified a defect in a small percentage of crimped wire connections in certain joysticks which could result in oxidation leading to the slowdown of wheelchair speed performance over time. In a very small percentage of instances, the power wheelchair may slow down relative to the commanded speed and then recover, creating an unintended acceleration. The unintended acceleration due to this defect has resulted in a consumer's power wheelchair running into objects, which can lead to bruising, pinched fingers, and pinched hands.

Invacare is providing distributors with two options to correct the affected joystick:

For details, please visit the following TGA website:
http://www.tga.gov.au/SARA/arn-detail.aspx?k=RC-2013-RN-00977-1

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 7 October 2013

Back