Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: ArjoHuntleigh Contoura 450 and Project Versions of 25000E/L Hospital Bed Frames

26 Nov 2013

The Medicines and Healthcare products Regulatory Agency (MHRA), United Kingdom (UK) posted a medical device safety alert concerning Contoura 450 and Project Versions of 25000E/L hospital bed frames manufactured by ArjoHuntleigh. Model numbers C450L, 25000E/LPD73, 25000E/LPD77, 25000E/LPD75, 25000E/LPD76, 25000E/LPD90 and 25000E/LPE12 are affected.

Spare backrest actuators (part number S6686, Linak ref. 321021) for the bed frames indicated above are supplied with a pre-wired safety limit switch (SLS) that operates as the actuator extends and prevents the backrest being driven to angles greater than 60o above the horizontal.

According to the manufacturer, approximately 200 of these actuators sold as spare parts were supplied with the SLS incorrectly connected. On beds fitted with these parts, if the backrest is driven to an angle of 60o, the SLS does not operate as intended and the backrest continues to rise until the actuator reaches its limit of travel. Once in this position, it may not be possible to lower the backrest, either electrically or by using the manual cardiopulmonary resuscitation (CPR) release. It is potentially life-threatening if the backrest cannot be lowered when a patient requires CPR or other emergency treatment.

ArjoHuntleigh advises the users to test all suspect actuators according to the instructions given in the field safety notice. Faulty actuators should be returned to ArjoHuntleigh for modification or disposal.

According to the local supplier, the affected products were not distributed in Hong Kong.

For details, please visit the following MHRA website:
http://www.mhra.gov.uk/Safetyinformation/Safetywarningsalertsandrecalls/fieldsafetynotices/CON336744

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 26 November 2013

Back