Safety Alerts

Safety Alerts

Starting from 1 January 2022, active safety alerts related to medical devices listed under MDACS are listed below. For safety alerts before 1 January 2022, please refer to archive.

Medical Device Safety Alert: Beckman Coulter Inc Coulter DxH Diluent

09 Jun 2023

Health Canada has posted a medical device safety alert concerning Beckman Coulter Inc Coulter DxH Diluent [Model or Catalogue Number: 628017] [HKMD No.: 200395].

According to the Health Canada, the manufacturer has confirmed that specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications. Quality control results may fail to flag for the affected parameters. Not all diluent containers within the affected lots exhibited the issue.

For further information, please visit https://recalls-rappels.canada.ca/en/alert-recall/coulter-dxh-diluent

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 9 Jun 2023

Back